Recall Z-0955-2020— Medtronic Inc.

Class I · Ongoing

Class I FDA medical device recall by Medtronic Inc., initiated 2019-11-21. FDA's recall record names no specific 510(k) clearance for this event. Product: Medtronic MiniMed 630G System with SmartGuard, Ref/Model # MMT-1515, MMT-1715, MMT-1714, MMT-1754, and MMT-1755 *US and Canada Version Only. Reason: There have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reservoir can become loose due to a broken or missing retainer ring that prevents a proper lock. If the reservoir is not properly locked into the pump, it could lead to over or under delivery of insulin, which could then result in hypoglycemia or hyperglycemia..

Recalling firm
Medtronic Inc.
Classification
Class I
Status
Ongoing
Initiated
2019-11-21
Firm location
Northridge, CA, United States

Product description

Medtronic MiniMed 630G System with SmartGuard, Ref/Model # MMT-1515, MMT-1715, MMT-1714, MMT-1754, and MMT-1755 *US and Canada Version Only

Reason for recall

There have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reservoir can become loose due to a broken or missing retainer ring that prevents a proper lock. If the reservoir is not properly locked into the pump, it could lead to over or under delivery of insulin, which could then result in hypoglycemia or hyperglycemia.

Data sourced from openFDA. This site is unofficial and independent of the FDA.