Recall Z-0952-2024— Philips North America
Class II · Ongoing
Class II FDA medical device recall by Philips North America, initiated 2024-01-03. It names one FDA 510(k) clearance: K193215. Product: Intera 1.5T Achieva Nova. Model (REF) Numbers 781172.. Reason: The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury..
- Recalling firm
- Philips North America
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-01-03
- Firm location
- Cambridge, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Intera 1.5T Achieva Nova. Model (REF) Numbers 781172.
Reason for recall
The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
Data sourced from openFDA. This site is unofficial and independent of the FDA.