Recall Z-0951-2020— Philips Respironics, Inc.
Class I · Terminated
Class I FDA medical device recall by Philips Respironics, Inc., initiated 2019-11-19, terminated 2021-03-31. It names one FDA 510(k) clearance: K181166. Product: Trilogy EVO Ventilator, 02 - Italy - Model Number: IT2100X21B Trilogy Evo is intended for pediatric through adult patients weighing at least 2.5 kg. The ventilator can measure, display, record, and alarm SpO2, FiO2, CO2, and Pulse Rate data when integrated with the appropriate accessories.. Reason: Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing requirements TSRS1298 and PRD439..
- Recalling firm
- Philips Respironics, Inc.
- Classification
- Class I
- Status
- Terminated
- Initiated
- 2019-11-19
- Terminated
- 2021-03-31
- Firm location
- Murrysville, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Trilogy EVO Ventilator, 02 - Italy - Model Number: IT2100X21B Trilogy Evo is intended for pediatric through adult patients weighing at least 2.5 kg. The ventilator can measure, display, record, and alarm SpO2, FiO2, CO2, and Pulse Rate data when integrated with the appropriate accessories.
Reason for recall
Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing requirements TSRS1298 and PRD439.
Data sourced from openFDA. This site is unofficial and independent of the FDA.