Recall Z-0950-2019— Arrow International Inc

Class II · Terminated

Class II FDA medical device recall by Arrow International Inc, initiated 2018-10-02, terminated 2021-11-12. It names one FDA 510(k) clearance: K090300. Product: AF 550 x 8ML/HR 2ML 15LO CPNBMT8015XL CP, Product code MT8015XL CPNB, Arrow AutoFuser Disposable Pain Control Pump. Reason: Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration..

Recalling firm
Arrow International Inc
Classification
Class II
Status
Terminated
Initiated
2018-10-02
Terminated
2021-11-12
Firm location
Reading, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

AF 550 x 8ML/HR 2ML 15LO CPNBMT8015XL CP, Product code MT8015XL CPNB, Arrow AutoFuser Disposable Pain Control Pump

Reason for recall

Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.

Data sourced from openFDA. This site is unofficial and independent of the FDA.