Recall Z-0945-2024— Philips North America

Class II · Ongoing

Class II FDA medical device recall by Philips North America, initiated 2024-01-03. It names one FDA 510(k) clearance: K193215. Product: Achieva 1.5T. Model (REF) Numbers 781196, 781343, 781296.. Reason: The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury..

Recalling firm
Philips North America
Classification
Class II
Status
Ongoing
Initiated
2024-01-03
Firm location
Cambridge, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Achieva 1.5T. Model (REF) Numbers 781196, 781343, 781296.

Reason for recall

The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.

Data sourced from openFDA. This site is unofficial and independent of the FDA.