Recall Z-0940-2021— Biomeme, Inc.

Class III · Terminated

Class III FDA medical device recall by Biomeme, Inc., initiated 2021-01-05, terminated 2021-06-11. FDA's recall record names no specific 510(k) clearance for this event. Product: Nucleic acid-based in vitro diagnostic devices for the detection of 2019-novel coronavirus (2019-nCoV) and/or diagnosis of 2019-nCoV infection in human clinical specimens. The RT-PCR test runs on a Franklin Real-Time PCR Thermocycler Device (Reference/Catalog Number 1000003,1000018) with the Software Controller version 6.4.3. The cycler has a "baseboard" unit with firmware and associated software component that control motorized movements of the optical filter carriage used for PCR detection.. Reason: Users cannot complete testing due to a sign-error in the software component that controls the filter movement and results in an instrument failure and assay failure before results are generated..

Recalling firm
Biomeme, Inc.
Classification
Class III
Status
Terminated
Initiated
2021-01-05
Terminated
2021-06-11
Firm location
Philadelphia, PA, United States

Product description

Nucleic acid-based in vitro diagnostic devices for the detection of 2019-novel coronavirus (2019-nCoV) and/or diagnosis of 2019-nCoV infection in human clinical specimens. The RT-PCR test runs on a Franklin Real-Time PCR Thermocycler Device (Reference/Catalog Number 1000003,1000018) with the Software Controller version 6.4.3. The cycler has a "baseboard" unit with firmware and associated software component that control motorized movements of the optical filter carriage used for PCR detection.

Reason for recall

Users cannot complete testing due to a sign-error in the software component that controls the filter movement and results in an instrument failure and assay failure before results are generated.

Data sourced from openFDA. This site is unofficial and independent of the FDA.