Recall Z-0934-2013— GE Healthcare, LLC
Class II · Terminated
Class II FDA medical device recall by GE Healthcare, LLC, initiated 2011-02-04, terminated 2014-01-27. It names 4 FDA 510(k) clearances: K092004, K061163, K052412 and 1 more. Product: GE Healthcare, Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ. Cardiovascular X-ray imaging systems. For use in generating fluoroscopic images of human anatomy.. Reason: Exposed high voltage wires inside the cabinet . On a cable located in the system cabinet, two of the shrink tubings covering the 400V live cable leave an area potentially unprotected..
- Recalling firm
- GE Healthcare, LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2011-02-04
- Terminated
- 2014-01-27
- Firm location
- Waukesha, WI, United States
510(k) clearances named in this recall
4 on FDA's record- K092004 — GE INNOVA SOLID STATE XRAY-IMAGER FLUOROSCOPIC X-RAY SYSTEM, MODEL: 4100IQ, 3100IQ, 2100IQ
- K061163 — INNOVA 2121IQ, INNOVA 3131IQ, INNOVA 4100, INNOVA 4100IQ, INNOVA 3100, INNOVA 3100IQ, INNOVA 2100IQ WITH INNOVA IVUS OPT
- K052412 — INNOVA 2100-IQ, INNOVA 3100, INNOVA 3100-IQ, INNOVA 4100, AND INNOVA 4100-IQ
- K050489 — INNOVA 2100
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
GE Healthcare, Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ. Cardiovascular X-ray imaging systems. For use in generating fluoroscopic images of human anatomy.
Reason for recall
Exposed high voltage wires inside the cabinet . On a cable located in the system cabinet, two of the shrink tubings covering the 400V live cable leave an area potentially unprotected.