Recall Z-0928-2018— Medtronic Navigation, Inc.-Littleton
Class II · Terminated
Class II FDA medical device recall by Medtronic Navigation, Inc.-Littleton, initiated 2018-01-25, terminated 2019-04-01. It names one FDA 510(k) clearance: K092564. Product: O-arm 1000 mobile image-intensified fluoroscopic x-ray system; (1) Model: BASE OARM Bl70000028100 SYS 100V, Product Number: Bl70000028100 (UDI: 00643169354418); (2) Model: BASE OARM Bl70000028120 SYS 120V, Product Number: Bl70000028120 (UDI: 00643169353411); (3) Model: BASE OARM Bl70000028120R SYS 120V RWK, Product Number: Bl70000028120R (UDI: 00643169353459); (4) Model: BASE OARM Bl70000028230 SYS 230V, Product Number: Bl70000028230 (UDI: 00643169353992); (5) BASE OARM Bl70000028230R SYS 230V RWK, Product Number: Bl70000028230R (UDI: 00643169354081) Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The O-arm Imaging System is compatible with certain Image Guided Surgery Systems.. Reason: Software update ("SW Update 3.2.1") is being implemented to address the following issues: Inability to power the system after shutdown, System stays in standalone mode, Image reconstruction, System/Pendant bootup, Dose display/report, Gantry motion, Network communication, System shutdown, and Early termination of 3D spin..
- Recalling firm
- Medtronic Navigation, Inc.-Littleton
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-01-25
- Terminated
- 2019-04-01
- Firm location
- Littleton, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
O-arm 1000 mobile image-intensified fluoroscopic x-ray system; (1) Model: BASE OARM Bl70000028100 SYS 100V, Product Number: Bl70000028100 (UDI: 00643169354418); (2) Model: BASE OARM Bl70000028120 SYS 120V, Product Number: Bl70000028120 (UDI: 00643169353411); (3) Model: BASE OARM Bl70000028120R SYS 120V RWK, Product Number: Bl70000028120R (UDI: 00643169353459); (4) Model: BASE OARM Bl70000028230 SYS 230V, Product Number: Bl70000028230 (UDI: 00643169353992); (5) BASE OARM Bl70000028230R SYS 230V RWK, Product Number: Bl70000028230R (UDI: 00643169354081) Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The O-arm Imaging System is compatible with certain Image Guided Surgery Systems.
Reason for recall
Software update ("SW Update 3.2.1") is being implemented to address the following issues: Inability to power the system after shutdown, System stays in standalone mode, Image reconstruction, System/Pendant bootup, Dose display/report, Gantry motion, Network communication, System shutdown, and Early termination of 3D spin.