Recall Z-0925-2020— Intuitive Surgical, Inc.

Class II · Terminated

Class II FDA medical device recall by Intuitive Surgical, Inc., initiated 2019-07-26, terminated 2022-08-10. It names one FDA 510(k) clearance: K131861. Product: da Vinci Xi Surgical System Surgical System, IS4000, 00886874114216 - Product Usage: used with an endoscope, surgical instruments, and accessories to perform minimally invasive surgical procedures.. Reason: The firm become aware of a potentially defective capacitors on circuit boards installed on certain da Vinci Xi, X and SP system Vision Side Carts (VSC) and/or Patient Side Carts (PSC). As a result, you may experience either vision loss or non-recoverable errors 307 or 319, preventing further use of the system..

Recalling firm
Intuitive Surgical, Inc.
Classification
Class II
Status
Terminated
Initiated
2019-07-26
Terminated
2022-08-10
Firm location
Sunnyvale, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

da Vinci Xi Surgical System Surgical System, IS4000, 00886874114216 - Product Usage: used with an endoscope, surgical instruments, and accessories to perform minimally invasive surgical procedures.

Reason for recall

The firm become aware of a potentially defective capacitors on circuit boards installed on certain da Vinci Xi, X and SP system Vision Side Carts (VSC) and/or Patient Side Carts (PSC). As a result, you may experience either vision loss or non-recoverable errors 307 or 319, preventing further use of the system.