Recall Z-0922-2018— Abbott Laboratories, Inc

Class II · Terminated

Class II FDA medical device recall by Abbott Laboratories, Inc, initiated 2017-12-08, terminated 2019-07-11. It names one FDA 510(k) clearance: K980367. Product: ARCHITECT c8000 Cuvette Segment, Part #01G46-01/01G06, a component to the ARCHITECT Clinical Chemistry system. There are 11 segments with each cuvette segment holding 15 cuvettes for a total of 165 cuvettes per reaction carousel. The firm name on the label is Abbott Laboratories, Abbott Park, IL. The Abbott ARCHITECT Clinical Chemistry Systems are designed to perform automated chemistry tests utilizing photometry and potentiometric technology. The cuvette segments are a component of the system and are racks that sit in the reaction carousel and hold cuvettes.. Reason: There is a potential to generate falsely depressed patient results in the cuvettes adjacent to the detached segment post for any assay run on the c4000, c8000, and c16000 instrument..

Recalling firm
Abbott Laboratories, Inc
Classification
Class II
Status
Terminated
Initiated
2017-12-08
Terminated
2019-07-11
Firm location
Irving, TX, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ARCHITECT c8000 Cuvette Segment, Part #01G46-01/01G06, a component to the ARCHITECT Clinical Chemistry system. There are 11 segments with each cuvette segment holding 15 cuvettes for a total of 165 cuvettes per reaction carousel. The firm name on the label is Abbott Laboratories, Abbott Park, IL. The Abbott ARCHITECT Clinical Chemistry Systems are designed to perform automated chemistry tests utilizing photometry and potentiometric technology. The cuvette segments are a component of the system and are racks that sit in the reaction carousel and hold cuvettes.

Reason for recall

There is a potential to generate falsely depressed patient results in the cuvettes adjacent to the detached segment post for any assay run on the c4000, c8000, and c16000 instrument.

Data sourced from openFDA. This site is unofficial and independent of the FDA.