Recall Z-0922-2017— Philips Electronics North America Corporation
Class II · Terminated
Class II FDA medical device recall by Philips Electronics North America Corporation, initiated 2016-12-01, terminated 2017-08-10. It names one FDA 510(k) clearance: K033737. Product: Allura Xper FD20 722012 is intended for: Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and Electro Physiology(EP).. Reason: Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled..
- Recalling firm
- Philips Electronics North America Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-12-01
- Terminated
- 2017-08-10
- Firm location
- Andover, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Allura Xper FD20 722012 is intended for: Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and Electro Physiology(EP).
Reason for recall
Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled.