Recall Z-0920-2020— Mindray DS USA, Inc. dba Mindray North America
Class II · Terminated
Class II FDA medical device recall by Mindray DS USA, Inc. dba Mindray North America, initiated 2019-12-09, terminated 2020-09-17. It names one FDA 510(k) clearance: K160370. Product: BA-800M Chemistry Analyzer Part Numbers BA81E-PA00002 and BA81E-PA00003, Containing Gas spring YQ-8/18-90-272-180N (Material Code M6T-010001--- (front spring)) and Gas spring YQ6/15-84-235(B-B)-80N (033-000090-00 (rear spring)). Reason: The gas spring on the front and rear protective cover may fail. There is a low risk of injury to the user resulting from failure of the gas spring. Injury may include impact to the arms, hands or fingers..
- Recalling firm
- Mindray DS USA, Inc. dba Mindray North America
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-12-09
- Terminated
- 2020-09-17
- Firm location
- Mahwah, NJ, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
BA-800M Chemistry Analyzer Part Numbers BA81E-PA00002 and BA81E-PA00003, Containing Gas spring YQ-8/18-90-272-180N (Material Code M6T-010001--- (front spring)) and Gas spring YQ6/15-84-235(B-B)-80N (033-000090-00 (rear spring))
Reason for recall
The gas spring on the front and rear protective cover may fail. There is a low risk of injury to the user resulting from failure of the gas spring. Injury may include impact to the arms, hands or fingers.
Data sourced from openFDA. This site is unofficial and independent of the FDA.