Recall Z-0905-2022— Stanbio Laboratory, LP

Class II · Terminated

Class II FDA medical device recall by Stanbio Laboratory, LP, initiated 2022-03-16, terminated 2025-01-02. FDA's recall record names no specific 510(k) clearance for this event. Product: Separation Technology Creamatocrit Plus, Catalog Number: 100-146. Reason: The device was being marketed as an IVD without a 510(k). Therefore, performance characteristics of these breast milk analyzers have not been adequately validated. An erroneous result could lead to sub-optimal nutrition supplementation for patients (infants).

Recalling firm
Stanbio Laboratory, LP
Classification
Class II
Status
Terminated
Initiated
2022-03-16
Terminated
2025-01-02
Firm location
Boerne, TX, United States

Product description

Separation Technology Creamatocrit Plus, Catalog Number: 100-146

Reason for recall

The device was being marketed as an IVD without a 510(k). Therefore, performance characteristics of these breast milk analyzers have not been adequately validated. An erroneous result could lead to sub-optimal nutrition supplementation for patients (infants)

Data sourced from openFDA. This site is unofficial and independent of the FDA.