Recall Z-0905-2020— Haemonetics Corporation

Class II · Terminated

Class II FDA medical device recall by Haemonetics Corporation, initiated 2019-10-21, terminated 2020-11-06. FDA's recall record names no specific 510(k) clearance for this event. Product: Japan DISP SET, 263J, 205J, 208 - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. Reason: Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing..

Recalling firm
Haemonetics Corporation
Classification
Class II
Status
Terminated
Initiated
2019-10-21
Terminated
2020-11-06
Firm location
Braintree, MA, United States

Product description

Japan DISP SET, 263J, 205J, 208 - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma

Reason for recall

Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.

Data sourced from openFDA. This site is unofficial and independent of the FDA.