Recall Z-0904-2022— Merit Medical Systems, Inc.
Class II · Ongoing
Class II FDA medical device recall by Merit Medical Systems, Inc., initiated 2022-03-01. FDA's recall record names no specific 510(k) clearance for this event. Product: AccessPLUS Large Bore Hemostasis Valves, Catalog Code: 580758001, 580774001, IN8152/C, K04-01197, K05T-01863, K09-09102B, K09-10388A, K09-12917, K09-13501AP, K10-05457AP, K10-05635, K10-05868P, K10-05872P, K12-08303, K12-10346, K12-10375, K12-10724AP, K12-10962, K12-11413, MAP150/F, MAP152/B, MAP152/F, K12-01551, K12-YP064A, K12-YP114, K12-YP115, K12-YS047A, IN8152, K12-01551. Reason: A design change made to a hemostasis valve to improve manufacturability caused an internal gap between the rotator and the Y-body, where the guidewire can get caught, resulting in difficulty advancing the guidewire through the device. This defect may result in a delay of procedure..
- Recalling firm
- Merit Medical Systems, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-03-01
- Firm location
- South Jordan, UT, United States
Product description
AccessPLUS Large Bore Hemostasis Valves, Catalog Code: 580758001, 580774001, IN8152/C, K04-01197, K05T-01863, K09-09102B, K09-10388A, K09-12917, K09-13501AP, K10-05457AP, K10-05635, K10-05868P, K10-05872P, K12-08303, K12-10346, K12-10375, K12-10724AP, K12-10962, K12-11413, MAP150/F, MAP152/B, MAP152/F, K12-01551, K12-YP064A, K12-YP114, K12-YP115, K12-YS047A, IN8152, K12-01551
Reason for recall
A design change made to a hemostasis valve to improve manufacturability caused an internal gap between the rotator and the Y-body, where the guidewire can get caught, resulting in difficulty advancing the guidewire through the device. This defect may result in a delay of procedure.
Data sourced from openFDA. This site is unofficial and independent of the FDA.