Recall Z-0899-2017— Radiometer America Inc

Class II · Terminated

Class II FDA medical device recall by Radiometer America Inc, initiated 2016-11-23, terminated 2017-02-07. It names 3 FDA 510(k) clearances: K151856, K142898, K130144. Product: AQURE System Software Version 2.2.0 Model #: 933-599 UDI: (01)05700699335999(10)2.2.0 Catalog number: 933-599 The AQURE system manages Radiometer blood gas and immunoassay analyzers and results from 3rd party devices for point of care testing placed throughout the hospital. It enables the user to track connected devices, monitor performance and availability, and to quickly resolve and document issues (including allowing the customer to track root cause and corrective action taken). Intended use: The AQURE system is intended to let you manage analytical devices and operator profiles. The user can associate patient data with test data. The system shows test results. The system receives data from connected devices at the point-of care or laboratory. It can send test results to the HIS/LIS. The system lets you send commands to selected devices. The system uses data related to the performance of devices, to tell users of issues to be managed. The AQURE system is intended for professional use.. Reason: Design error when displaying additional information in the patient view window; error may result in misreading a parameter and its value..

Recalling firm
Radiometer America Inc
Classification
Class II
Status
Terminated
Initiated
2016-11-23
Terminated
2017-02-07
Firm location
Brea, CA, United States

510(k) clearances named in this recall

3 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

AQURE System Software Version 2.2.0 Model #: 933-599 UDI: (01)05700699335999(10)2.2.0 Catalog number: 933-599 The AQURE system manages Radiometer blood gas and immunoassay analyzers and results from 3rd party devices for point of care testing placed throughout the hospital. It enables the user to track connected devices, monitor performance and availability, and to quickly resolve and document issues (including allowing the customer to track root cause and corrective action taken). Intended use: The AQURE system is intended to let you manage analytical devices and operator profiles. The user can associate patient data with test data. The system shows test results. The system receives data from connected devices at the point-of care or laboratory. It can send test results to the HIS/LIS. The system lets you send commands to selected devices. The system uses data related to the performance of devices, to tell users of issues to be managed. The AQURE system is intended for professional use.

Reason for recall

Design error when displaying additional information in the patient view window; error may result in misreading a parameter and its value.