Recall Z-0896-2022— DRG International, Inc.
Class II · Terminated
Class II FDA medical device recall by DRG International, Inc., initiated 2021-12-21, terminated 2024-02-12. FDA's recall record names no specific 510(k) clearance for this event. Product: C-Peptide ELISA : IVD measurement of C-Peptide in serum, plasma (EDTA-, heparin- or citrate plasma) and urine. Catalog # EIA-1293. Reason: ODs of the standards too low(< 0.6 ODmax) may cause invalid runs, delayed patient results are possible.
- Recalling firm
- DRG International, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2021-12-21
- Terminated
- 2024-02-12
- Firm location
- Springfield, NJ, United States
Product description
C-Peptide ELISA : IVD measurement of C-Peptide in serum, plasma (EDTA-, heparin- or citrate plasma) and urine. Catalog # EIA-1293
Reason for recall
ODs of the standards too low(< 0.6 ODmax) may cause invalid runs, delayed patient results are possible
Data sourced from openFDA. This site is unofficial and independent of the FDA.