Recall Z-0893-2015— Roche Diagnostics Operations, Inc.

Class II · Terminated

Class II FDA medical device recall by Roche Diagnostics Operations, Inc., initiated 2014-11-20, terminated 2018-01-10. It names one FDA 510(k) clearance: K972135. Product: ACCU-CHEK Tender II 17/80 10 + 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections through the injection port. The indwelling catheter can also be securely attached to the infusion catheter by means of a proprietary click-lock connector for the infusion of drugs subcutaneously. The infusion set can be detached from the indwelling catheter, and the catheter capped to allow freedom from the infusion set and pump for showers, athletics or other activities.. Reason: Roche Diabetes Care was informed by UnoMedical, the Manufacturer of ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets, about potential for the the tubing of the infusion set to become detached at the connect/disconnect location on the ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets. If tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify.

Recalling firm
Roche Diagnostics Operations, Inc.
Classification
Class II
Status
Terminated
Initiated
2014-11-20
Terminated
2018-01-10
Firm location
Indianapolis, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ACCU-CHEK Tender II 17/80 10 + 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections through the injection port. The indwelling catheter can also be securely attached to the infusion catheter by means of a proprietary click-lock connector for the infusion of drugs subcutaneously. The infusion set can be detached from the indwelling catheter, and the catheter capped to allow freedom from the infusion set and pump for showers, athletics or other activities.

Reason for recall

Roche Diabetes Care was informed by UnoMedical, the Manufacturer of ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets, about potential for the the tubing of the infusion set to become detached at the connect/disconnect location on the ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets. If tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify

Data sourced from openFDA. This site is unofficial and independent of the FDA.