Recall Z-0887-2026— Bard Peripheral Vascular Inc

Class II · Ongoing

Class II FDA medical device recall by Bard Peripheral Vascular Inc, initiated 2025-11-06. FDA's recall record names no specific 510(k) clearance for this event. Product: Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,. Reason: Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels..

Recalling firm
Bard Peripheral Vascular Inc
Classification
Class II
Status
Ongoing
Initiated
2025-11-06
Firm location
Tempe, AZ, United States

Product description

Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,

Reason for recall

Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels.