Recall Z-0885-2016— Biomet, Inc.

Class II · Terminated

Class II FDA medical device recall by Biomet, Inc., initiated 2016-02-16, terminated 2017-02-14. FDA's recall record names no specific 510(k) clearance for this event. Product: HipLOC CHS Plate/Lag Screw Introducer. Reason: A complaint was received where the Hiploc Plate/Lag Screw Introducer was undersized and would not accept the introducor coupling screw. It appears now that not all parts manufactured to Rev A drawing were reworked..

Recalling firm
Biomet, Inc.
Classification
Class II
Status
Terminated
Initiated
2016-02-16
Terminated
2017-02-14
Firm location
Warsaw, IN, United States

Product description

HipLOC CHS Plate/Lag Screw Introducer

Reason for recall

A complaint was received where the Hiploc Plate/Lag Screw Introducer was undersized and would not accept the introducor coupling screw. It appears now that not all parts manufactured to Rev A drawing were reworked.