Recall Z-0884-2013— Philips Healthcare Inc.

Class II · Terminated

Class II FDA medical device recall by Philips Healthcare Inc., initiated 2012-10-25, terminated 2014-03-31. It names one FDA 510(k) clearance: K982795. Product: Philips DigitalDiagnost X-Ray System with Eleva Software 2.1.3 Model numbers 712020, 712022, and 712082 Product Usage: Digital radiographic system. Reason: When the operator for a wall stand view selects an "image rotation" different from default, or such image rotation is preset in the examination database, then the resulting image may be rotated in the wrong direction. When the operator manually rotates the image back, electronic side markers (if programmed) are rotated also and as a result may be placed in a wrong position inside the image..

Recalling firm
Philips Healthcare Inc.
Classification
Class II
Status
Terminated
Initiated
2012-10-25
Terminated
2014-03-31
Firm location
Andover, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Philips DigitalDiagnost X-Ray System with Eleva Software 2.1.3 Model numbers 712020, 712022, and 712082 Product Usage: Digital radiographic system

Reason for recall

When the operator for a wall stand view selects an "image rotation" different from default, or such image rotation is preset in the examination database, then the resulting image may be rotated in the wrong direction. When the operator manually rotates the image back, electronic side markers (if programmed) are rotated also and as a result may be placed in a wrong position inside the image.