Recall Z-0882-2013— Philips Medical Systems (Cleveland) Inc

Class II · Terminated

Class II FDA medical device recall by Philips Medical Systems (Cleveland) Inc, initiated 2012-10-24, terminated 2015-07-23. It names one FDA 510(k) clearance: K052640. Product: Philips Medical Systems, GEMINI LXL - Model #882412; GEMINI TF 16 - Model #882470 and Model #882473; GEMINI TF 64 - Model #882471, Positron Emission Tomography (PET) and X-Ray Computed Tomography (CT) scanning systems, diagnostic imaging systems for fixed or mobile installations that combine Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.. Reason: Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm whereby these generated datasets are 'flipped' and incorrectly labeled left-to-right when the are saved to disk. If the flipped Intensity Project.

Recalling firm
Philips Medical Systems (Cleveland) Inc
Classification
Class II
Status
Terminated
Initiated
2012-10-24
Terminated
2015-07-23
Firm location
Cleveland, OH, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Philips Medical Systems, GEMINI LXL - Model #882412; GEMINI TF 16 - Model #882470 and Model #882473; GEMINI TF 64 - Model #882471, Positron Emission Tomography (PET) and X-Ray Computed Tomography (CT) scanning systems, diagnostic imaging systems for fixed or mobile installations that combine Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.

Reason for recall

Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm whereby these generated datasets are 'flipped' and incorrectly labeled left-to-right when the are saved to disk. If the flipped Intensity Project

Data sourced from openFDA. This site is unofficial and independent of the FDA.