Recall Z-0876-2018— Bausch & Lomb Inc Irb

Class II · Terminated

Class II FDA medical device recall by Bausch & Lomb Inc Irb, initiated 2017-03-03, terminated 2018-06-25. FDA's recall record names no specific 510(k) clearance for this event. Product: LASEREDGE Knives, 2.8MM SLIT KNIFE ANGLED DOUBLE BEVEL (6/BOX), REF/PRODUCT CODE E7552ADB, STERILE, Rx Only Product Usage: The LaserEdge Knives are intended for ophthalmic surgery. There is no therapeutic use and they are available to surgeons for use in procedures.. Reason: The complaints database was reviewed and there have been an atypical volume of dull blades complaints for various lots. out of an abundance of precaution, Bausch + Lomb is voluntarily recalling all product codes SKU manufacturing without the electro-polish control and penetration force testing..

Recalling firm
Bausch & Lomb Inc Irb
Classification
Class II
Status
Terminated
Initiated
2017-03-03
Terminated
2018-06-25
Firm location
Rochester, NY, United States

Product description

LASEREDGE Knives, 2.8MM SLIT KNIFE ANGLED DOUBLE BEVEL (6/BOX), REF/PRODUCT CODE E7552ADB, STERILE, Rx Only Product Usage: The LaserEdge Knives are intended for ophthalmic surgery. There is no therapeutic use and they are available to surgeons for use in procedures.

Reason for recall

The complaints database was reviewed and there have been an atypical volume of dull blades complaints for various lots. out of an abundance of precaution, Bausch + Lomb is voluntarily recalling all product codes SKU manufacturing without the electro-polish control and penetration force testing.

Data sourced from openFDA. This site is unofficial and independent of the FDA.