Recall Z-0874-2019— Philips Medical Systems Nederlands
Class II · Terminated
Class II FDA medical device recall by Philips Medical Systems Nederlands, initiated 2018-12-21, terminated 2020-11-19. It names one FDA 510(k) clearance: K030520. Product: Intera 0.5T Standard, Model Number 781101. Nuclear Magnetic Resonance Imaging System. Reason: There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field..
- Recalling firm
- Philips Medical Systems Nederlands
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-12-21
- Terminated
- 2020-11-19
- Firm location
- Best, Netherlands
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Intera 0.5T Standard, Model Number 781101. Nuclear Magnetic Resonance Imaging System
Reason for recall
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Data sourced from openFDA. This site is unofficial and independent of the FDA.