Recall Z-0874-2015— Ebi, Llc
Class II · Terminated
Class II FDA medical device recall by Ebi, Llc, initiated 2014-11-11, terminated 2015-04-15. It names one FDA 510(k) clearance: K113796. Product: Solitaire-C Cervical Spacer System a) 14-531593 Drill/Awl Sleeve b) 14-531594 Spring-Loaded Drill/Awl Sleeve Intended to stabilize cervical spinal segment to promote fusion in order to restrict motion and decrease pain using bone graft with or without supplemental fixation.. Reason: A design deficiency was discovered whereby the Drill/Awl Sleeve and Spring-Loaded Drill/Awl Sleeve instruments (the Sleeves) allow the mating Drill and Awl instruments to penetrate 1.0 mm beyond their intended depths during screw hole preparation..
- Recalling firm
- Ebi, Llc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-11-11
- Terminated
- 2015-04-15
- Firm location
- Parsippany, NJ, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Solitaire-C Cervical Spacer System a) 14-531593 Drill/Awl Sleeve b) 14-531594 Spring-Loaded Drill/Awl Sleeve Intended to stabilize cervical spinal segment to promote fusion in order to restrict motion and decrease pain using bone graft with or without supplemental fixation.
Reason for recall
A design deficiency was discovered whereby the Drill/Awl Sleeve and Spring-Loaded Drill/Awl Sleeve instruments (the Sleeves) allow the mating Drill and Awl instruments to penetrate 1.0 mm beyond their intended depths during screw hole preparation.