Recall Z-0872-2013— Hill-Rom, Inc.

Class II · Terminated

Class II FDA medical device recall by Hill-Rom, Inc., initiated 2012-10-31, terminated 2016-03-09. FDA's recall record names no specific 510(k) clearance for this event. Product: Liko Overhead Rail System, Ceiling Mounted Rail Systems for all Liko Stationary Lifts including Likorall, Multirall and Masterlift Systems Product Usage - The Overhead Rail System allows overhead lifts to be mounted to the ceiling allowing versatility for use in lifting situations. It can be used with different Liko ceiling lifts.. Reason: Liko/Hill-Rom to date has received two reports from facilities alleging that the overhead rail system has failed to the extent that the overhead rails have fallen down while in use in high humidity (swimming pool) areas. Analysis of the incidents suggest that the components (hardware) used to secure the rail to the ceiling had corroded and may be unsuitable for environments with an atmosphere con.

Recalling firm
Hill-Rom, Inc.
Classification
Class II
Status
Terminated
Initiated
2012-10-31
Terminated
2016-03-09
Firm location
Batesville, IN, United States

Product description

Liko Overhead Rail System, Ceiling Mounted Rail Systems for all Liko Stationary Lifts including Likorall, Multirall and Masterlift Systems Product Usage - The Overhead Rail System allows overhead lifts to be mounted to the ceiling allowing versatility for use in lifting situations. It can be used with different Liko ceiling lifts.

Reason for recall

Liko/Hill-Rom to date has received two reports from facilities alleging that the overhead rail system has failed to the extent that the overhead rails have fallen down while in use in high humidity (swimming pool) areas. Analysis of the incidents suggest that the components (hardware) used to secure the rail to the ceiling had corroded and may be unsuitable for environments with an atmosphere con

Data sourced from openFDA. This site is unofficial and independent of the FDA.