Recall Z-0870-2020— Energetiq Technology Inc

Class II · Ongoing

Class II FDA medical device recall by Energetiq Technology Inc, initiated 2019-09-27. FDA's recall record names no specific 510(k) clearance for this event. Product: EQ-400-RH-QZ and EQ-400-LH-BK products Product Usage: The device is a very broadband light source emitting incoherent non-laser light from the Deep UV through the visible wavelengths.. Reason: On September 27, 2019 Energetiq Technology (the manufacturer) discovered a problem with their EQ-400-RH-QZ and EQ-400-LH-BK products during routine manufacturing functional testing. It was noted that a safety interlock within the lamp does not function in a reliable manner to shut down the system if the laser collimator is removed from the lamp while the system is operating. This is a potential safety hazard for anyone working on or nearby the lamp..

Recalling firm
Energetiq Technology Inc
Classification
Class II
Status
Ongoing
Initiated
2019-09-27
Firm location
Woburn, MA, United States

Product description

EQ-400-RH-QZ and EQ-400-LH-BK products Product Usage: The device is a very broadband light source emitting incoherent non-laser light from the Deep UV through the visible wavelengths.

Reason for recall

On September 27, 2019 Energetiq Technology (the manufacturer) discovered a problem with their EQ-400-RH-QZ and EQ-400-LH-BK products during routine manufacturing functional testing. It was noted that a safety interlock within the lamp does not function in a reliable manner to shut down the system if the laser collimator is removed from the lamp while the system is operating. This is a potential safety hazard for anyone working on or nearby the lamp.

Data sourced from openFDA. This site is unofficial and independent of the FDA.