Recall Z-0869-2020— LivaNova USA Inc
Class II · Terminated
Class II FDA medical device recall by LivaNova USA Inc, initiated 2018-11-16, terminated 2022-12-12. FDA's recall record names no specific 510(k) clearance for this event. Product: Cyberonics VNS Therapy AspireSR, Model 106 Generator, 26-0007-8400/1. The firm name on the label is CYBERONICS, INC., Houston, TX. Vagus nerve stimulator.. Reason: This recall is an expansion of Z-3019-2017 and Z-3020-2017, which was initiated to fix the devices premature battery depletion, caused by electrical leakage on the circuit board assemblies of the Models 105 and 106 generators. *Note this recall occurred in November 2018..
- Recalling firm
- LivaNova USA Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-11-16
- Terminated
- 2022-12-12
- Firm location
- Houston, TX, United States
Product description
Cyberonics VNS Therapy AspireSR, Model 106 Generator, 26-0007-8400/1. The firm name on the label is CYBERONICS, INC., Houston, TX. Vagus nerve stimulator.
Reason for recall
This recall is an expansion of Z-3019-2017 and Z-3020-2017, which was initiated to fix the devices premature battery depletion, caused by electrical leakage on the circuit board assemblies of the Models 105 and 106 generators. *Note this recall occurred in November 2018.