Recall Z-0865-2017— Accuray Incorporated

Class II · Terminated

Class II FDA medical device recall by Accuray Incorporated, initiated 2016-12-06, terminated 2017-06-09. It names one FDA 510(k) clearance: K011024. Product: CyberKnife Robotic Radiosurgery Systems, models: G3, G4, VSI. Radiology: The CyberKnife Robotic Radiosurgery System is indicated for treatment planning and image guided stereotactic Radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.. Reason: Accuray has become aware of a potential safety issue involving the possibility of electric shock during maintenance activities from the lon Pump HV Power Supply at the back of the modulator cabinet. This issue is limited to certain G3, G4, and VSI models of the CyberKnife¿ Robotic Radiosurgery System..

Recalling firm
Accuray Incorporated
Classification
Class II
Status
Terminated
Initiated
2016-12-06
Terminated
2017-06-09
Firm location
Sunnyvale, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

CyberKnife Robotic Radiosurgery Systems, models: G3, G4, VSI. Radiology: The CyberKnife Robotic Radiosurgery System is indicated for treatment planning and image guided stereotactic Radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Reason for recall

Accuray has become aware of a potential safety issue involving the possibility of electric shock during maintenance activities from the lon Pump HV Power Supply at the back of the modulator cabinet. This issue is limited to certain G3, G4, and VSI models of the CyberKnife¿ Robotic Radiosurgery System.