Recall Z-0864-2021— Philips Ultrasound Inc
Class II · Terminated
Class II FDA medical device recall by Philips Ultrasound Inc, initiated 2020-10-27, terminated 2023-04-19. It names one FDA 510(k) clearance: K182857. Product: All EPIQ and Affiniti Ultrasound systems. Model numbers - EPIQ 5G, EPIC 5C, EPIQ 5W, EPIQ 7G, EPIC 7C, EPIQ 7W, EPIQ CVx, Affiniti 30, Affiniti 50, Affiniti 70. Diagnostic Ultrasound System. MOD - EPIQ 5 GMDN - 40761 100-240V 50/60 Hz, 600 VA MOD - EPIQ 7 GMDN - 40761 100-240V 50/60 Hz, 600 VA MOD - EPIQ 100-240V 50/60 Hz, 600 VA MOD - Affinity 30 GMDN - 40761 100-240V 50/60 Hz, 450 VA MOD - Affinity 50 GMDN - 40761 100-240V 50/60 Hz, 450 VA MOD - Affinity 70 GMDN - 40761 100-240V 50/60 Hz, 450 VA. Reason: The manufacturer has determined that with certain uncommon workflows there is potential for incorrect patient data to be displayed and saved into an exam. The issue(s) manifest differently for different versions of software..
- Recalling firm
- Philips Ultrasound Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-10-27
- Terminated
- 2023-04-19
- Firm location
- Bothell, WA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
All EPIQ and Affiniti Ultrasound systems. Model numbers - EPIQ 5G, EPIC 5C, EPIQ 5W, EPIQ 7G, EPIC 7C, EPIQ 7W, EPIQ CVx, Affiniti 30, Affiniti 50, Affiniti 70. Diagnostic Ultrasound System. MOD - EPIQ 5 GMDN - 40761 100-240V 50/60 Hz, 600 VA MOD - EPIQ 7 GMDN - 40761 100-240V 50/60 Hz, 600 VA MOD - EPIQ 100-240V 50/60 Hz, 600 VA MOD - Affinity 30 GMDN - 40761 100-240V 50/60 Hz, 450 VA MOD - Affinity 50 GMDN - 40761 100-240V 50/60 Hz, 450 VA MOD - Affinity 70 GMDN - 40761 100-240V 50/60 Hz, 450 VA
Reason for recall
The manufacturer has determined that with certain uncommon workflows there is potential for incorrect patient data to be displayed and saved into an exam. The issue(s) manifest differently for different versions of software.