Recall Z-0864-2016— EKOS Corporation

Class II · Terminated

Class II FDA medical device recall by EKOS Corporation, initiated 2016-01-21, terminated 2016-03-15. It names 3 FDA 510(k) clearances: K140151, K111705, K081467. Product: EkoSonic Endovascular System, Catalog Number 500-56112. It consists of three main components, a single use EkoSonic Endovascular Device which is comprised of an Intelligent Drug Delivery Catheter (IDDC) and removable MicroSonic Device (MSD), and a reusable EKOS EkoSonic Control System. During use, fluids are delivered through the side-holes of the IDDC. The MSD is placed through the central lumen of the IDDC to deliver ultrasound energy along the length of fluid delivery, dispersing the fluid into the intravascular treatment site. The reusable EKOS EkoSonic Control Unit provides power to the system and the user interface for operator control. A reusable, non-sterile Connector Interface Cable connects the EKOS EkoSonic Control Unit to the EkoSonic Endovascular Device.. Reason: The EkoSonic kit, including the individually labeled MicroSonic Device (MSD) and Intelligent Drug Delivery Catheter (IDDC) are labeled with a 135 cm working length, and programmed with a 135 cm family code, however the MSD actual working length is 106 cm..

Recalling firm
EKOS Corporation
Classification
Class II
Status
Terminated
Initiated
2016-01-21
Terminated
2016-03-15
Firm location
Bothell, WA, United States

510(k) clearances named in this recall

3 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

EkoSonic Endovascular System, Catalog Number 500-56112. It consists of three main components, a single use EkoSonic Endovascular Device which is comprised of an Intelligent Drug Delivery Catheter (IDDC) and removable MicroSonic Device (MSD), and a reusable EKOS EkoSonic Control System. During use, fluids are delivered through the side-holes of the IDDC. The MSD is placed through the central lumen of the IDDC to deliver ultrasound energy along the length of fluid delivery, dispersing the fluid into the intravascular treatment site. The reusable EKOS EkoSonic Control Unit provides power to the system and the user interface for operator control. A reusable, non-sterile Connector Interface Cable connects the EKOS EkoSonic Control Unit to the EkoSonic Endovascular Device.

Reason for recall

The EkoSonic kit, including the individually labeled MicroSonic Device (MSD) and Intelligent Drug Delivery Catheter (IDDC) are labeled with a 135 cm working length, and programmed with a 135 cm family code, however the MSD actual working length is 106 cm.