Recall Z-0863-2019— Physio-Control Inc
Class I · Terminated
Class I FDA medical device recall by Physio-Control Inc, initiated 2019-02-01, terminated 2024-04-18. It names one FDA 510(k) clearance: K142430. Product: LIFEPAK 15 Monitor/Defibrillator service kits Product Usage: The LIFEPAK(R) 15 Monitor/defibrillator (LP15) is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. The LP15 monitor/defibrillator is intended for use by trained medical personnel in out-of-doors and indoor emergency care settings within the environmental conditions specified in the Operating Instructions. The LP15 monitor/defibrillator is designed to be used during ground transportation except when specified otherwise. Manual mode monitoring and therapy functions are intended for use on adult and pediatric patients. Automated external defibrillation mode is intended for use on patients eight years of age and older.. Reason: Certain LIFEPAK 15 Monitors/ Defibrillators were reported to experience a lockup condition after a shock was delivered. This condition is defined as a blank monitor display with LED lights on, indicating power on the device, but no response in keypad and device functions..
- Recalling firm
- Physio-Control Inc
- Classification
- Class I
- Status
- Terminated
- Initiated
- 2019-02-01
- Terminated
- 2024-04-18
- Firm location
- Redmond, WA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
LIFEPAK 15 Monitor/Defibrillator service kits Product Usage: The LIFEPAK(R) 15 Monitor/defibrillator (LP15) is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. The LP15 monitor/defibrillator is intended for use by trained medical personnel in out-of-doors and indoor emergency care settings within the environmental conditions specified in the Operating Instructions. The LP15 monitor/defibrillator is designed to be used during ground transportation except when specified otherwise. Manual mode monitoring and therapy functions are intended for use on adult and pediatric patients. Automated external defibrillation mode is intended for use on patients eight years of age and older.
Reason for recall
Certain LIFEPAK 15 Monitors/ Defibrillators were reported to experience a lockup condition after a shock was delivered. This condition is defined as a blank monitor display with LED lights on, indicating power on the device, but no response in keypad and device functions.