Recall Z-0862-2020— Siemens Healthcare Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2019-10-11, terminated 2021-05-03. It names one FDA 510(k) clearance: K151767. Product: Atellica CH 930 Analyzer, Siemens Material Number 11067000, Software Versions V1.19.2 or 1.20.0 The Atellica CH 930 Analyzer is an automated, clinical chemistry analyzer designed to perform in vitro diagnostic tests on clinical specimens.. Reason: On the Atellica CH 930 Analyzer, when scanning barcodes at the module console for Integrated Multisensor Technology (IMT) system fluids (Standard A (Std A) and Standard B (Std B)) for Sodium (Na), Potassium (K), and Chloride (Cl), the system will translate the decimal point separators to commas. This will result in invalid concentration values for the fluids, therefore the system will utilize default nominal values instead of lot specific IMT fluid concentration values for calculation of the calibrations and results for the Na, K, and Cl methods..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2019-10-11
Terminated
2021-05-03
Firm location
Tarrytown, NY, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Atellica CH 930 Analyzer, Siemens Material Number 11067000, Software Versions V1.19.2 or 1.20.0 The Atellica CH 930 Analyzer is an automated, clinical chemistry analyzer designed to perform in vitro diagnostic tests on clinical specimens.

Reason for recall

On the Atellica CH 930 Analyzer, when scanning barcodes at the module console for Integrated Multisensor Technology (IMT) system fluids (Standard A (Std A) and Standard B (Std B)) for Sodium (Na), Potassium (K), and Chloride (Cl), the system will translate the decimal point separators to commas. This will result in invalid concentration values for the fluids, therefore the system will utilize default nominal values instead of lot specific IMT fluid concentration values for calculation of the calibrations and results for the Na, K, and Cl methods.