Recall Z-0853-2014— GE Healthcare, LLC

Class II · Terminated

Class II FDA medical device recall by GE Healthcare, LLC, initiated 2013-11-15, terminated 2014-03-28. It names one FDA 510(k) clearance: K122457. Product: GE Healthcare Innova IGS 630. Biplane Cardiovascular and Interventional Imaging System. Product Usage: The angiographic X-ray systems are indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures. Additionally, with the OR table, the angiographic X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for image-guided surgical procedures. The OR table is suitable for interventional and surgical procedures. The Cathiab Frontiers solutions are indicated for use in conjunction with single plane and biplane GE angiographic X-ray systems and imaging / data medical devices used in interventional and surgical cathlab environments and cleared for commercial distribution. The Cathiab Frontiers solutions are integrated GE angiographic X-ray and imaging / data medical devices that simplify the end-to-end clinical workflow in the cathlab by implementing: 1) Communication protocols for exchanging and automatically synchronizing patient, exam, system, and image information between the angiographic X-ray systems and the imaging / data medical devices. 2) Communication protocols for the control of imaging / data medical device functions from the angiographic X-ray systems user interface. 3) Interfaces for displaying the imaging / data medical device output on the monitor display solutions of the GE angiographic X-ray systems.. Reason: GE Healthcare has recently become aware of a potential safety issue with respect to IGS 630 Imaging Systems. While performing fluoroscopy on a biplane fluoroscopy unit, there is a potential for loss of the x-ray imaging function when the user changes the size of the lateral FOV (Field of View) and releases the biplane footswitch pedal simultaneously. This could lead to a system lock up requiring.

Recalling firm
GE Healthcare, LLC
Classification
Class II
Status
Terminated
Initiated
2013-11-15
Terminated
2014-03-28
Firm location
Waukesha, WI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

GE Healthcare Innova IGS 630. Biplane Cardiovascular and Interventional Imaging System. Product Usage: The angiographic X-ray systems are indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures. Additionally, with the OR table, the angiographic X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for image-guided surgical procedures. The OR table is suitable for interventional and surgical procedures. The Cathiab Frontiers solutions are indicated for use in conjunction with single plane and biplane GE angiographic X-ray systems and imaging / data medical devices used in interventional and surgical cathlab environments and cleared for commercial distribution. The Cathiab Frontiers solutions are integrated GE angiographic X-ray and imaging / data medical devices that simplify the end-to-end clinical workflow in the cathlab by implementing: 1) Communication protocols for exchanging and automatically synchronizing patient, exam, system, and image information between the angiographic X-ray systems and the imaging / data medical devices. 2) Communication protocols for the control of imaging / data medical device functions from the angiographic X-ray systems user interface. 3) Interfaces for displaying the imaging / data medical device output on the monitor display solutions of the GE angiographic X-ray systems.

Reason for recall

GE Healthcare has recently become aware of a potential safety issue with respect to IGS 630 Imaging Systems. While performing fluoroscopy on a biplane fluoroscopy unit, there is a potential for loss of the x-ray imaging function when the user changes the size of the lateral FOV (Field of View) and releases the biplane footswitch pedal simultaneously. This could lead to a system lock up requiring