Recall Z-0848-2022— RAYSEARCH LABORATORIES AB

Class II · Terminated

Class II FDA medical device recall by RAYSEARCH LABORATORIES AB, initiated 2021-09-09, terminated 2024-04-16. It names one FDA 510(k) clearance: K210645. Product: TomoHelical and TomoDirect plans in RayStation 6, 7, 8A, 8B, 9A, 9B, 10A, 10B and 11A. Build Numbers: 6.1.1.2, 6.2.0.7, 6.3.0.6, 7.0.0.19, 8.0.0.61, 8.0.1.10, 8.1.0.47, 8.1.1.8, 8.1.2.5, 9.0.0.113, 9.1.0.933, 9.2.0.483, 10.0.0.1154, 10.0.1.52, 10.1.0.613, 11.0.0.951 or 11.0.1.29.. Reason: If the dynamic jaw mode is used and an optimization is continued after changing the dose grid or modifying the target ROI, the positions of the jaws may change unexpectedly for some control points. User must be aware to avoid unnecessary jaw openings during treatment planning.

Recalling firm
RAYSEARCH LABORATORIES AB
Classification
Class II
Status
Terminated
Initiated
2021-09-09
Terminated
2024-04-16
Firm location
Stockholm, N/A, Sweden

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

TomoHelical and TomoDirect plans in RayStation 6, 7, 8A, 8B, 9A, 9B, 10A, 10B and 11A. Build Numbers: 6.1.1.2, 6.2.0.7, 6.3.0.6, 7.0.0.19, 8.0.0.61, 8.0.1.10, 8.1.0.47, 8.1.1.8, 8.1.2.5, 9.0.0.113, 9.1.0.933, 9.2.0.483, 10.0.0.1154, 10.0.1.52, 10.1.0.613, 11.0.0.951 or 11.0.1.29.

Reason for recall

If the dynamic jaw mode is used and an optimization is continued after changing the dose grid or modifying the target ROI, the positions of the jaws may change unexpectedly for some control points. User must be aware to avoid unnecessary jaw openings during treatment planning