Recall Z-0848-2016— Zimmer Biomet, Inc.

Class II · Terminated

Class II FDA medical device recall by Zimmer Biomet, Inc., initiated 2016-01-11, terminated 2017-05-23. It names one FDA 510(k) clearance: K973714. Product: VERSYS 6 INCH BEADED FC 1 prosthesis, hip, semi-constrained, metal/polymer, porous uncemented "" Total hip replacement for the following: severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, nonunion of previous fractures of the femur; congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis; disability due to previous fusion; previously failed endoprostheses, and/or total hip components in the affected extremity and acute femoral neck fractures. " Hemi-hip replacement for the following: fracture dislocation of the hip; elderly, debilitated patients when a total hip replacement is contraindicated; irreducible fractures in which adequate fixation cannot be obtained; certain high subcapital fractures and comminuted femoral neck fractures in the aged; nonunion of femoral neck fractures; secondary avascular necrosis of the femoral head; pathological fractures of the femoral neck; and osteoarthritis in which the femoral head is primarily affected.". Reason: LDPE bag containing the implant adheres to the highly polished implant surface..

Recalling firm
Zimmer Biomet, Inc.
Classification
Class II
Status
Terminated
Initiated
2016-01-11
Terminated
2017-05-23
Firm location
Warsaw, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

VERSYS 6 INCH BEADED FC 1 prosthesis, hip, semi-constrained, metal/polymer, porous uncemented "" Total hip replacement for the following: severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, nonunion of previous fractures of the femur; congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis; disability due to previous fusion; previously failed endoprostheses, and/or total hip components in the affected extremity and acute femoral neck fractures. " Hemi-hip replacement for the following: fracture dislocation of the hip; elderly, debilitated patients when a total hip replacement is contraindicated; irreducible fractures in which adequate fixation cannot be obtained; certain high subcapital fractures and comminuted femoral neck fractures in the aged; nonunion of femoral neck fractures; secondary avascular necrosis of the femoral head; pathological fractures of the femoral neck; and osteoarthritis in which the femoral head is primarily affected."

Reason for recall

LDPE bag containing the implant adheres to the highly polished implant surface.

Data sourced from openFDA. This site is unofficial and independent of the FDA.