Recall Z-0842-2014— Biomerieux France Chemin De L'Or

Class II · Terminated

Class II FDA medical device recall by Biomerieux France Chemin De L'Or, initiated 2013-11-08, terminated 2017-11-30. FDA's recall record names no specific 510(k) clearance for this event. Product: API Listeria (REF 10300) API Listeria is a standardized system for the identification of Listeria which uses miniaturized tests, as well as a database. The ZYM B reagent is including in API Listeria kit and it is used as an additional reagent for revealing the result of DIM miniaturized biochemical test included in the API Listeria strips. ZYM B reagent is added in the DIM test before reading miniaturized biochemical test including in the API Listeria strips.. Reason: bioM¿rieux identified a visual defect and activity issue on the ZYM B reagent (REF 70493); leading to a false negative results of some API biochemical tests using the ZYM B reagent..

Recalling firm
Biomerieux France Chemin De L'Or
Classification
Class II
Status
Terminated
Initiated
2013-11-08
Terminated
2017-11-30
Firm location
Marcy L'Etoile, France

Product description

API Listeria (REF 10300) API Listeria is a standardized system for the identification of Listeria which uses miniaturized tests, as well as a database. The ZYM B reagent is including in API Listeria kit and it is used as an additional reagent for revealing the result of DIM miniaturized biochemical test included in the API Listeria strips. ZYM B reagent is added in the DIM test before reading miniaturized biochemical test including in the API Listeria strips.

Reason for recall

bioM¿rieux identified a visual defect and activity issue on the ZYM B reagent (REF 70493); leading to a false negative results of some API biochemical tests using the ZYM B reagent.

Data sourced from openFDA. This site is unofficial and independent of the FDA.