Recall Z-0841-2015— GE OEC Medical Systems, Inc
Class II · Terminated
Class II FDA medical device recall by GE OEC Medical Systems, Inc, initiated 2014-09-15, terminated 2015-05-20. It names one FDA 510(k) clearance: K123603. Product: GE OEC Brivo 865 Plus Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize.. Reason: mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3)..
- Recalling firm
- GE OEC Medical Systems, Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-09-15
- Terminated
- 2015-05-20
- Firm location
- Salt Lake City, UT, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
GE OEC Brivo 865 Plus Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize.
Reason for recall
mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).