Recall Z-0840-2015— GE OEC Medical Systems, Inc
Class II · Terminated
Class II FDA medical device recall by GE OEC Medical Systems, Inc, initiated 2014-09-15, terminated 2015-05-20. It names one FDA 510(k) clearance: K123603. Product: GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize.. Reason: mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3)..
- Recalling firm
- GE OEC Medical Systems, Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-09-15
- Terminated
- 2015-05-20
- Firm location
- Salt Lake City, UT, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize.
Reason for recall
mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).
Data sourced from openFDA. This site is unofficial and independent of the FDA.