Recall Z-0837-2014— DeRoyal Industries Inc

Class II · Terminated

Class II FDA medical device recall by DeRoyal Industries Inc, initiated 2013-09-17, terminated 2017-06-05. FDA's recall record names no specific 510(k) clearance for this event. Product: DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee Blanket w/ Straps NS, REF T654NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.. Reason: The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock..

Recalling firm
DeRoyal Industries Inc
Classification
Class II
Status
Terminated
Initiated
2013-09-17
Terminated
2017-06-05
Firm location
New Tazewell, TN, United States

Product description

DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee Blanket w/ Straps NS, REF T654NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.

Reason for recall

The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.