Recall Z-0834-2013— Ventana Medical Systems Inc

Class II · Terminated

Class II FDA medical device recall by Ventana Medical Systems Inc, initiated 2012-09-28, terminated 2016-02-24. FDA's recall record names no specific 510(k) clearance for this event. Product: VENTANA (a member of the Roche Group) DISCOVERY ULTRA; REF/MU 750-601; Research Use Only Product Usage: Research Use Only-The DISCOVERY ULTRA Advanced Staining System is intended to automatically stain histological or cytological specimens on microscope slides with specific inununohistochemistry or in situ hybridization reagents.. Reason: Potential for a waste fluid overflow condition to accur while running decontamination cycles on the BenchMark ULTRA automated staining platform or the DISCOVERY ULTRA automated staining platform.The waste fluid drainage system may become restricted by build-up of debris in the tubing and in-line filter, thereby inhibiting optimal flow rates. This type of overflow event can lead to migration of flu.

Recalling firm
Ventana Medical Systems Inc
Classification
Class II
Status
Terminated
Initiated
2012-09-28
Terminated
2016-02-24
Firm location
Oro Valley, AZ, United States

Product description

VENTANA (a member of the Roche Group) DISCOVERY ULTRA; REF/MU 750-601; Research Use Only Product Usage: Research Use Only-The DISCOVERY ULTRA Advanced Staining System is intended to automatically stain histological or cytological specimens on microscope slides with specific inununohistochemistry or in situ hybridization reagents.

Reason for recall

Potential for a waste fluid overflow condition to accur while running decontamination cycles on the BenchMark ULTRA automated staining platform or the DISCOVERY ULTRA automated staining platform.The waste fluid drainage system may become restricted by build-up of debris in the tubing and in-line filter, thereby inhibiting optimal flow rates. This type of overflow event can lead to migration of flu

Data sourced from openFDA. This site is unofficial and independent of the FDA.