Recall Z-0833-2014— Symmetry Medical/SSI
Class II · Terminated
Class II FDA medical device recall by Symmetry Medical/SSI, initiated 2013-12-05, terminated 2015-04-13. It names one FDA 510(k) clearance: K120117. Product: Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container. Sterilization container systems manufactured and distributed under the following names: Small Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Small Perforated Bottom, REF 50-8731, 270mm x 270mm x 100mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Small Perforated Bottom, REF 50-8732, 270mm x 270mm x 135mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Small Perforated Bottom, REF 50-8733, 270mm x 270mm x 150mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Small Perforated Bottom, REF 50-8734, 270mm x 270mm x 200mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Small Perforated Bottom, REF 50-8735, 270mm x 270mm x 260mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Small Non-Perforated Bottom, REF 50-8744, 270mm x 270mm x 100mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Small Non-Perforated Bottom, REF 50-8745, 270mm x 270mm x 135mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Small Non-Perforated Bottom, REF 50-8746, 270mm x 270mm x 150mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Small Non-Perforated Bottom, REF 50-8747, 270mm x 270mm x 200mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Small Non-Perforated Bottom, REF 50-8748, 270mm x 270mm x 260mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Lid Grey Small 4 Latch, REF 50-8950, 280mm x 276mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Lid Yellow Small 4 Latch, REF 50-8951, 280mm x 276mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Lid Green Small 4 Latch, REF 50-8952, 280mm x 276mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Lid Blue Small 4 Latch, REF 50-8953, 280mm x 276mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Lid Red Small 4 Latch, REF 50-8954, 280mm x 276mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Lid Black Small 4 Latch, REF 50-8955, 280mm x 276mm, QTY: 1, Symmetry Surgical, Inc., Medium Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Medium Perforated Bottom, REF 50-8736, 440mm x 270mm x 100mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Medium Perforated Bottom, REF 50-8737, 440mm x 270mm x 135mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Medium Perforated Bottom, REF 50-8738, 440mm x 270mm x 150mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Medium Non-Perforated Bottom, REF 50-8749, 440mm x 270mm x 100mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Medium Non-Perforated Bottom, REF 50-8750, 440mm x 270mm x 135mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Medium Non-Perforated Bottom, REF 50-8751, 440mm x 270mm x 150mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Lid Grey Medium 4 Latch, REF 50-8956, 445mm x 279mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Lid Yellow Medium 4 Latch, REF 50-8957, 445mm x 279mm, QTY: 1,. Reason: The Quad-Lock Sterilization Container System may not maintain an effective sterile barrier..
- Recalling firm
- Symmetry Medical/SSI
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-12-05
- Terminated
- 2015-04-13
- Firm location
- Antioch, TN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container. Sterilization container systems manufactured and distributed under the following names: Small Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Small Perforated Bottom, REF 50-8731, 270mm x 270mm x 100mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Small Perforated Bottom, REF 50-8732, 270mm x 270mm x 135mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Small Perforated Bottom, REF 50-8733, 270mm x 270mm x 150mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Small Perforated Bottom, REF 50-8734, 270mm x 270mm x 200mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Small Perforated Bottom, REF 50-8735, 270mm x 270mm x 260mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Small Non-Perforated Bottom, REF 50-8744, 270mm x 270mm x 100mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Small Non-Perforated Bottom, REF 50-8745, 270mm x 270mm x 135mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Small Non-Perforated Bottom, REF 50-8746, 270mm x 270mm x 150mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Small Non-Perforated Bottom, REF 50-8747, 270mm x 270mm x 200mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Small Non-Perforated Bottom, REF 50-8748, 270mm x 270mm x 260mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Lid Grey Small 4 Latch, REF 50-8950, 280mm x 276mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Lid Yellow Small 4 Latch, REF 50-8951, 280mm x 276mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Lid Green Small 4 Latch, REF 50-8952, 280mm x 276mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Lid Blue Small 4 Latch, REF 50-8953, 280mm x 276mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Lid Red Small 4 Latch, REF 50-8954, 280mm x 276mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Lid Black Small 4 Latch, REF 50-8955, 280mm x 276mm, QTY: 1, Symmetry Surgical, Inc., Medium Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Medium Perforated Bottom, REF 50-8736, 440mm x 270mm x 100mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Medium Perforated Bottom, REF 50-8737, 440mm x 270mm x 135mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Medium Perforated Bottom, REF 50-8738, 440mm x 270mm x 150mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Medium Non-Perforated Bottom, REF 50-8749, 440mm x 270mm x 100mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Medium Non-Perforated Bottom, REF 50-8750, 440mm x 270mm x 135mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Medium Non-Perforated Bottom, REF 50-8751, 440mm x 270mm x 150mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Lid Grey Medium 4 Latch, REF 50-8956, 445mm x 279mm, QTY: 1, Symmetry Surgical, Inc., Symmetry and CODMAN brands, QUAD-LOCK Sterilization Container Lid Yellow Medium 4 Latch, REF 50-8957, 445mm x 279mm, QTY: 1,
Reason for recall
The Quad-Lock Sterilization Container System may not maintain an effective sterile barrier.
Data sourced from openFDA. This site is unofficial and independent of the FDA.