Recall Z-0832-2024— American Contract Systems, Inc.

Class II · Ongoing

Class II FDA medical device recall by American Contract Systems, Inc., initiated 2023-11-29. FDA's recall record names no specific 510(k) clearance for this event. Product: Custom procedural convenience kits and trays, Ophthalmic use, labeled as: a) EYE PACK, kit number LVEY27; b) Laminectomy, kit number UIOL34AG. Reason: During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals..

Recalling firm
American Contract Systems, Inc.
Classification
Class II
Status
Ongoing
Initiated
2023-11-29
Firm location
Temple Terrace, FL, United States

Product description

Custom procedural convenience kits and trays, Ophthalmic use, labeled as: a) EYE PACK, kit number LVEY27; b) Laminectomy, kit number UIOL34AG

Reason for recall

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

Data sourced from openFDA. This site is unofficial and independent of the FDA.