Recall Z-0831-2021— Fresenius Medical Care Holdings, Inc.
Class II · Ongoing
Class II FDA medical device recall by Fresenius Medical Care Holdings, Inc., initiated 2020-11-23. It names one FDA 510(k) clearance: K070177. Product: Naturalyte Liquid Acid concentrate bottles, Product Code 08-1251-1 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.. Reason: The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification..
- Recalling firm
- Fresenius Medical Care Holdings, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2020-11-23
- Firm location
- Waltham, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Naturalyte Liquid Acid concentrate bottles, Product Code 08-1251-1 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
Reason for recall
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
Data sourced from openFDA. This site is unofficial and independent of the FDA.