Class II FDA medical device recall by Baxter Healthcare Corporation, initiated 2022-02-21. FDA's recall record names no specific 510(k) clearance for this event. Product: Liko Universal Twinbar 670 QRH, an accessory that can be used together with most Liko lifts, model numbers 3156087 and P3156087.. Reason: An incorrect center bolt was used in the assembly of the product which could result in a patient fall during use..

Recalling firm
Baxter Healthcare Corporation
Classification
Class II
Status
Ongoing
Initiated
2022-02-21
Firm location
Deerfield, IL, United States

Product description

Liko Universal Twinbar 670 QRH, an accessory that can be used together with most Liko lifts, model numbers 3156087 and P3156087.

Reason for recall

An incorrect center bolt was used in the assembly of the product which could result in a patient fall during use.