Recall Z-0825-2025— Boston Scientific Corporation

Class I · Ongoing

Class I FDA medical device recall by Boston Scientific Corporation, initiated 2024-12-12. FDA's recall record names no specific 510(k) clearance for this event. Product: ALTRUA 2 DR Pacemaker, SL (Model Number S702) and EL (Model Number S722). Reason: A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance..

Recalling firm
Boston Scientific Corporation
Classification
Class I
Status
Ongoing
Initiated
2024-12-12
Firm location
Saint Paul, MN, United States

Product description

ALTRUA 2 DR Pacemaker, SL (Model Number S702) and EL (Model Number S722)

Reason for recall

A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.