Recall Z-0825-2014— Siemens Medical Solutions USA, Inc.
Class II · Terminated
Class II FDA medical device recall by Siemens Medical Solutions USA, Inc., initiated 2013-12-22, terminated 2015-03-05. It names one FDA 510(k) clearance: K123622. Product: ACUSON SC2000 Power Input: 100-240VAC, 1600 Max 50/60Hz; Siemens Medical Solutions USA, Inc. 685 East Middlefield Road Mountain View, CA 94043 Ultrasound imaging system.. Reason: The locking mechanism that is intended to hold the control panel in a fixed position can become loose and fail to lock the rotation of the control panel, making it unable to steer or control during transport..
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-12-22
- Terminated
- 2015-03-05
- Firm location
- Mountain View, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
ACUSON SC2000 Power Input: 100-240VAC, 1600 Max 50/60Hz; Siemens Medical Solutions USA, Inc. 685 East Middlefield Road Mountain View, CA 94043 Ultrasound imaging system.
Reason for recall
The locking mechanism that is intended to hold the control panel in a fixed position can become loose and fail to lock the rotation of the control panel, making it unable to steer or control during transport.
Data sourced from openFDA. This site is unofficial and independent of the FDA.