Recall Z-0824-2025— Boston Scientific Corporation

Class I · Ongoing

Class I FDA medical device recall by Boston Scientific Corporation, initiated 2024-12-12. FDA's recall record names no specific 510(k) clearance for this event. Product: ESSENTIO DR Pacemaker, SL (Model Number L101, SL MRI (Model Number L111), EL (Model Number L121), and EL MRI (Model Number L131). Reason: A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance..

Recalling firm
Boston Scientific Corporation
Classification
Class I
Status
Ongoing
Initiated
2024-12-12
Firm location
Saint Paul, MN, United States

Product description

ESSENTIO DR Pacemaker, SL (Model Number L101, SL MRI (Model Number L111), EL (Model Number L121), and EL MRI (Model Number L131)

Reason for recall

A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.

Data sourced from openFDA. This site is unofficial and independent of the FDA.