Recall Z-0823-2014— Siemens Healthcare Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2013-10-30, terminated 2014-12-15. FDA's recall record names no specific 510(k) clearance for this event. Product: Siemens CentraLink" Data Management System V14.0.4, V14.0.5 or V14.0.8 The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location.. Reason: Under extremely rare circumstances, a patient result that had been previously disabled may be released without proper review. (Disabled Test in CentraLink System or Aptio System May Become Enabled in CentraLink System.).

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2013-10-30
Terminated
2014-12-15
Firm location
Newark, DE, United States

Product description

Siemens CentraLink" Data Management System V14.0.4, V14.0.5 or V14.0.8 The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location.

Reason for recall

Under extremely rare circumstances, a patient result that had been previously disabled may be released without proper review. (Disabled Test in CentraLink System or Aptio System May Become Enabled in CentraLink System.)

Data sourced from openFDA. This site is unofficial and independent of the FDA.