Recall Z-0822-2025— Boston Scientific Corporation
Class I · Ongoing
Class I FDA medical device recall by Boston Scientific Corporation, initiated 2024-12-12. FDA's recall record names no specific 510(k) clearance for this event. Product: ACCOLADE DR Pacemaker, SL (Model Number L301), SL MRI (Model Number L311), EL (Model Number L321), and EL MRI (Model Number L331). Reason: A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance..
- Recalling firm
- Boston Scientific Corporation
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2024-12-12
- Firm location
- Saint Paul, MN, United States
Product description
ACCOLADE DR Pacemaker, SL (Model Number L301), SL MRI (Model Number L311), EL (Model Number L321), and EL MRI (Model Number L331)
Reason for recall
A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
Data sourced from openFDA. This site is unofficial and independent of the FDA.